Climate chamber Uniprotect
Code CCU
The air flow cabinet provides homogenous air distribution across all levels within the cabinet. It is fitted with a HEPA filter system for excellent air conditions.
Apskatīt citus produktus pēc pielietojuma
funkcijos
Key features
- Homogenous air distribution across all levels within the cabinet
- HEPA filter system H 13, according to DIN EN 1822, for supply and exhaust air with fi lter monitoring and device for DEHS test
(optional with HEPA fi lter system H 14) - Outer casing made of stainless steel / AISI 304 powder coated, chamber made of stainless steel / AISI 304
- Easy to clean and to disinfect
Uniprotect design
- 7“ TFT Colour touch panel with integrated control for
- easy operation
- Two energy-saving blowers with operation monitoring
- Double door with door monitoring
- Pressure control of chamber at negative and positive
- pressure
- LED illumination with alternating light (red/white) via
- Daily / weekly timer
- Air vent
- Display air exchange (fresh air supply)
- Connection to the building’s heating / air conditioning
Unirpotect equipment
- Shelves 4 shelves (thereof 3 adjustable) of 1440 x 675 mm useful space
- Adjusting range 32 mm
- Load capacity / shelve 30 kg
- Mobility 4 castors (thereof 2 lockable), adjustable leg levelers
- Double doors, basquill lock, barrier-free access
- Windows safety glass, 5mm thick
- Prefilter supply air G4, as per DIN EN 779; 570x270x14 mm
- Prefilter exhaust air G4, as per DIN EN 779; 640x260x14 mm
- HEPA filter supply air H14, as per DIN EN 1822; 305x610x78 mm
- HEPA filter exhaust air H14, as per DIN EN 1822; 305x610x78 mm
- Illumination LED white/red, programmable, dimmable
- HVAC exhaust socket, outer diameter Ø ca. 114 mm
- To be connected in combination with a draught damper
Electronic properties
- Electrical connection 230 V / 50 Hz | 1/N/PE/AC, max. 16-amps-backup
- Power consumption 0,2 kW (with active heating 1,1 kW)
- Electrical protection class I
- Electrical protection IP 40
- Current 4,8 A
- Ports RJ 45, USB
- BMS Plug Round plug, 6-pin
- Control features Signal lamps: Operation ON | Info| Alarm |
- Touchpanel/HMI, touch sensitive screen
- Technical properties
- Temperature range 5°C above ambient, with optional heating (up to a max. 35°C)
- Max. temperature 50°C
- Control unit SPS-control, positive/negative pressure, switchable, adjustable
- Air flow / noise level horizontal, left to right / < 53 dB[A]
- Differential pressure sensors -500 to +1.000 Pa -> 0 – 10 V or 4 – 20 mA (supply and exhaust air)
- Filter monitoring Differential pressure control 50 – 500 Pa (default setting 250 Pa)
- Flow rate monitoring Air velocity sensor 0 – 10 m/s
Cage capacities for UniProtect climate chamber
- Type II 24
- Type II L 24
- Type III 20
- Type IV 12
Specifikacija
Dimensions
- Outside (WxHxD) mm 1551 x 1800 x 721 (without door handles)
- With humidifier 1743 x 1808 x 798
- Inside (WxHxD) mm 1440 x 1250 x 675, stainless steel material no. 1.4301/AISI 304
- Capacity 1.210 liters
- Weight approx. 331 kg net / approx. 380 kg gross
Komponentai
Standarta sistēmas stastāvdaļas
- Optional operating under positive or negative pressure
- Red windows
- Temperature control
- Filter monitoring
- Device for DEHS measurement
Sistēmas papildinājumi
- Active humidification up to 60 % rH
- Active carbon filter for odor reduction
- Draught damper for connection to the HVAC